LIS

Administrative Code

Creating a Report: Check the sections you'd like to appear in the report, then use the "Create Report" button at the bottom of the page to generate your report. Once the report is generated you'll then have the option to download it as a pdf, print or email the report.

Virginia Administrative Code
Title 12. Health
Agency 5. Department of Health
Chapter 481. Virginia Radiation Protection Regulations
11/21/2024

12VAC5-481-2045. Full calibration measurements.

A. Teletherapy units.

1. Licensees authorized to use a teletherapy unit for medical use shall perform full calibration measurements on each teletherapy unit:

a. Before the first medical use of the unit;

b. Before medical use under the following conditions:

(1) Whenever spot-check measurements indicate that the output differs by more than 5.0% from the output obtained at the last full calibration corrected mathematically for radioactive decay;

(2) Following replacement of the source or following reinstallation of the teletherapy unit in a new location; and

(3) Following any repair of the teletherapy unit that includes removal of the source or major repair of the components associated with the source exposure assembly; and

c. At intervals not exceeding one year.

2. To satisfy the requirement of subdivision 1 of this subsection, full calibration measurements shall include determination of:

a. The output within plus or minus 3.0% for the range of field sizes and for the distance or range of distances used for medical use;

b. The coincidence of the radiation field and the field indicated by the light beam localizing device;

c. The uniformity of the radiation field and its dependence on the orientation of the useful beam;

d. Timer accuracy and linearity over the range of use;

e. On-off error; and

f. The accuracy of all distance measuring and localization devices in medical use.

3. Licensees shall use the dosimetry system described in 12VAC5-481-2044 to measure the output for one set of exposure conditions. The remaining radiation measurements required in subdivision 2 a of this subsection may be made using a dosimetry system that indicates relative dose rates.

4. Licensees shall make full calibration measurements required by subdivision 1 of this subsection in accordance with published protocols accepted by nationally recognized bodies.

5. Licensees shall mathematically correct the outputs determined in subdivision 2 a of this subsection for physical decay for intervals not exceeding one month for cobalt-60, six months for cesium-137, or at intervals consistent with 1.0% decay for all other nuclides.

6. Full calibration measurements required by subdivision 1 of this subsection and physical decay corrections required by subdivision 5 of this subsection shall be performed by the authorized medical physicist (AMP).

B. Remote afterloader units.

1. Licensees authorized to use a remote afterloader unit for medical use shall perform full calibration measurements on each unit:

a. Before the first medical use of the unit;

b. Before medical use under the following conditions:

(1) Following replacement of the source or following reinstallation of the unit in a new location outside the facility; and

(2) Following any repair of the unit that includes removal of the source or major repair of the components associated with the source exposure assembly;

c. At intervals not exceeding one quarter for high dose-rate, medium dose-rate, and pulsed dose-rate remote afterloader units with sources whose half-life exceeds 75 days; and

d. At intervals not exceeding one year for low dose-rate remote afterloader units.

2. To satisfy the requirement of subdivision 1 of this subsection, full calibration measurements shall include, as applicable, determination of:

a. The output within plus or minus 5.0%;

b. Source positioning accuracy to within plus or minus 1 millimeter;

c. Source retraction with backup battery upon power failure;

d. Length of the source transfer tubes;

e. Timer accuracy and linearity over the typical range of use;

f. Length of the applicators; and

g. Function of the source transfer tubes, applicators, and transfer tube-applicator interfaces.

3. Licensees shall use the dosimetry system described in 12VAC5-481-2044 to measure the output.

4. Licensees shall make full calibration measurements required by subdivision 1 of this subsection in accordance with published protocols accepted by nationally recognized bodies.

5. In addition to the requirements for full calibrations for low dose-rate remote afterloader units in subdivision 2 of this subsection, licensees shall perform an autoradiograph of the sources to verify inventory and source arrangement at intervals not exceeding one calendar quarter.

6. For low dose-rate remote afterloader units, licensees may use measurements provided by the source manufacturer that are made in accordance with subdivisions 1 through 5 of this subsection.

7. Licensees shall mathematically correct the outputs determined in subdivision 2 a of this subsection for physical decay at intervals consistent with 1.0% physical decay.

8. Full calibration measurements required by subdivision 1 of this subsection and physical decay corrections required by subdivision 7 of this subsection shall be performed by the AMP.

C. Gamma stereotactic radiosurgery units.

1. Licensees authorized to use a gamma stereotactic radiosurgery unit for medical use shall perform full calibration measurements on each unit:

a. Before the first medical use of the unit;

b. Before medical use under the following conditions:

(1) Whenever spot-check measurements indicate that the output differs by more than 5.0% from the output obtained at the last full calibration corrected mathematically for radioactive decay;

(2) Following replacement of the sources or following reinstallation of the gamma stereotactic radiosurgery unit in a new location; and

(3) Following any repair of the gamma stereotactic radiosurgery unit that includes removal of the sources or major repair of the components associated with the source assembly; and

c. At intervals not exceeding one year, with the exception that relative helmet factors need only be determined before the first medical use of a helmet and following any damage to a helmet.

2. To satisfy the requirement of subdivision 1 of this subsection, full calibration measurements shall include determination of:

a. The output within plus or minus 3.0%;

b. Relative helmet factors;

c. Isocenter coincidence;

d. Timer accuracy and linearity over the range of use;

e. On-off error;

f. Trunnion centricity;

g. Treatment table retraction mechanism, using backup battery power or hydraulic backups with the unit off;

h. Helmet microswitches;

i. Emergency timing circuits; and

j. Stereotactic frames and localizing devices (trunnions).

3. Licensees shall use the dosimetry system described in 12VAC5-481-2044 to measure the output for one set of exposure conditions. The remaining radiation measurements required in subdivision 2 a of this subsection may be made using a dosimetry system that indicates relative dose rates.

4. Licensees shall make full calibration measurements required by subdivision 1 of this subsection in accordance with published protocols accepted by nationally recognized bodies.

5. Licensees shall mathematically correct the outputs determined in subdivision 2 a of this subsection at intervals not exceeding one month for cobalt-60 and at intervals consistent with 1.0% physical decay for all other radionuclides.

6. Full calibration measurements required by subdivision 1 of this subsection and physical decay corrections required by subdivision 5 of this subsection shall be performed by the AMP.

Statutory Authority

§ 32.1-229 of the Code of Virginia.

Historical Notes

Derived from Virginia Register Volume 32, Issue 24, eff. August 25, 2016.

Website addresses provided in the Virginia Administrative Code to documents incorporated by reference are for the reader's convenience only, may not necessarily be active or current, and should not be relied upon. To ensure the information incorporated by reference is accurate, the reader is encouraged to use the source document described in the regulation.

As a service to the public, the Virginia Administrative Code is provided online by the Virginia General Assembly. We are unable to answer legal questions or respond to requests for legal advice, including application of law to specific fact. To understand and protect your legal rights, you should consult an attorney.